Company record
AGFA Healthcare Corp.
0 FDA warning letters since 2023 and 1 recall in the last 12 months.
Get an email the day FDA posts anything new about AGFA Healthcare Corp..
Recalls
| Reported | Class | Firm | Product and reason |
|---|---|---|---|
| Dec 31, 2025 | Class II | AGFA Healthcare Corp. Device |
Brand name: AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic Common Name: DR 800 It is possible to exceed the maximum allowed Air Kerma rate (AKR) of 88mGy/min for pulsed fluoro exams. This can happen in case the framerate is increased on the NX workstation when ABS is not enabled. FDA notice Z-0923-2026 |